产品中心
Product Center
当前位置:首页
产品中心
日本菊水kikusui仪表
日本菊水kikusui│泄漏电流测试仪
TOS1200泄漏电流测试仪日本菊水TOS1200泄漏电流测试仪
                    
产品简介
                  Calibration of Leakage Current Detection Sensitivity; Direct Reading of Error from Error Display Scale; Ammeter Ranges; Eliminates Need for Power Supply; AC/DC Selection Switch;
TOS1200泄漏电流测试仪
Calibration of Leakage Current Detection Sensitivity
Direct Reading of Error from Error Display Scale
Ammeter Ranges
Eliminates Need for Power Supply
AC/DC Selection Switch
| Measuring Function | Measurement of current values and error (%) for AC (50/60 Hz) and DC at a test voltage of 1000 V | 
| Measuring Ranges | 8 ranges consisting of 0.5/1/2/5/10/20/50/100 mA along with values equal to 0.8 times the values of those ranges (for 1, 2, 4 and 8 steps) | 
| Ammeter Scale | Main scale: Direct-reading errordisplay scale over a range of ±10%of the above full scale values Auxiliary scale: Ratio scale of 0 to1.1 times the above full scale valuesAccuracy: (equivalent to 0% display of mainscale when the ratio is equal to 1) 
  |         
| Ammeter Accuracy | Main scale: ±1% of reading Auxiliary scale: ±3% of full scale value | 
| Ammeter Indication | DC/AC(sine wave rms value calibration of mean value response) | 
| Load Resistance | |||
| Range[mA] | Resistance[kΩ] | Range[mA] | Resistance[kΩ] | 
| 0.5 | 2000 | 10 | 100 | 
| 1 | 1000 | 20 | 50 | 
| 2 | 500 | 50 | 20 | 
| 5 | 200 | 100 | 10 | 
| Allowed Input Time | 0.5/1/2/5 mA ranges: Continuous 10/20/50/100 mA ranges: 60 sec. Max. 1/3 of duty cycle  |         ||
| Dimensions (MAX) | 134W × 164H × 270D mm (140W × 189H × 320D mm)  |         ||
| Weight | approx. 3.5 kg | ||
| Accessories | TL04-TOS high-voltage test lead: 1 | ||